Oxygen Regulator Bull Nose, Dual Output
Description:
Economy Medical Oxygen regulator – bull nose with dual output to Patient via Barb and to a device via Female Heyer.
Intended Use:
To reduce the pressure from the medical oxygen cylinder and to control the flow rate of medical oxygen to the patient. (Medical device to be used by a medical professional or under their instruction).
Features:
The flow selection dial has 12 positions, including 11 discrete flow rates and an off (0) position Flow is indicated in Liters per minute (ℓ/min) and is visible through a window adjacent to the flow selection dial. Cylinder contents gauge for convenient verification of cylinder contents. Rubber bumper guard around gauge. Dual output – Female Heyer with constant pressure and barb outlet with variable flow. A service exchange option is available should the regulator get damaged and the female Heyer is still in good working order. Easy installation and operation.
Technical Specifications:
Gas: Medical Oxygen
Material: Aluminium Body, Brass Chamber
Input Connection: Bull nose – Medical Oxygen
Input Pressure: ≤ 15100 kPa (151 Bar or 2190 psi)
Output Connection: Dual outlet:
Output 1: Barb outlet – Variable flow
Output 2: Female Heyer outlet – Constant pressure
Output Pressure: 414 kPa (4.14 bar or 60 psi) ± 5%
Output 1 Flow Setting: 0, 0.5, 1, 2, 3, 4, 5, 6, 8, 10, 15, and 25 ℓ/min
Output 1 Flow Accuracy: 0 to 1 ℓ/min – Accuracy: ±20% 2 to 25 ℓ/min – Accuracy: ±15%
Output 2 Maximum Flow: ≤50 ℓ/min
(NOTE: It is not advised to use the device at maximum flow rate for extended periods, as the regulator might be subject to cooling and might freeze up.)