Oxygen (O₂) Regulator Bullnose Dual Outlet

Oxygen (O₂) Regulator Bullnose Dual Outlet

Barbed Male & DISS / Ventilator Fitting (not for use with bubble through humidifier)

Oxygen (O₂) Regulator Bullnose Dual Outlet

Description:

Oxygen (O₂) Regulator Bullnose Dual Outlet with patient and device output.

Intended Use:

To reduce the pressure from the medical oxygen cylinder to administer oxygen to the patient via output 1, and to supply devices via output 2. (Medical device to be used by a medical professional or under their instruction).

Features:

The flow selection dial has 12 positions, including 11 discrete flow rates and an off (0) position Flow is indicated in litres per minute (ℓ/min) and is visible through a window adjacent to the flow selection dial. Cylinder contents gauge for convenient verification of cylinder contents. Rubber bumper guard around gauge. Dual output. Easy installation and operation.

Technical Specifications:

Gas: Medical Oxygen

Material: Aluminium Body, Brass Chamber

Input Connection: Bull Nose – Medical Oxygen

Input Pressure: ≤ 15100 kPa (151 Bar or 2190 psi)

Dual Output Connections:  Output 1: Barb outlet – Variable flow applications (for patient connection)

Output 2: 14.3 mm (9 /16 in) Ø outlet – Constant pressure applications (for devices)

Output Pressure: 414 kPa (4.14 bar or 60 psi) ± 5%

Output 1 Flow Setting: 0, 0.5, 1, 2, 3, 4, 5, 6, 8, 10, 15, and 25 ℓ/min

Output 1 Flow Accuracy:  0 to 1 ℓ/min – Accuracy: ±20%

2 to 25 ℓ/min – Accuracy: ±15%

Output 2 Maximum Flow: ≤50 ℓ/min

(NOTE: It is not advised to use the device at maximum flow rate for extended periods, as the regulator might be subject to cooling and might freeze up.)

Disclaimer: Specifications may change without prior notice, all images shown are for illustrative purposes only. Actual product may vary.

 